Termostatik fırın, çok fonksiyonlu kalibratör ve dijital iletişim cihazı
Temperature generation (at 23°C): -30/+150°C
• Thermal stability: 0.01°C
• Heating time from -30 to +150°C: 23 minutes
• Cooling time from 150 to 30°C: 37 minutes
• Extended temperature uncertainty with internal sensor: up to 0.15°C
• Measurement and generation of voltage, current and resistance signals
• Measurement and generation of frequencies and pulses
• Complete test switch
• Integrated measurement loop supply
• Measurement and simulation of thermocouple signals (13 types) and temperature resistors (14 types)
• Possibility to customize temperature resistors with the insertion of the Callendar Van correction coefficients
Dusen
• Possibility to connect an unlimited number of interchangeable external pressure modules: measuring ranges
selectable between +/- 10 mbar and 0-1000 bar, with accuracies up to +/- 0.005% f.s. + 0.01% reading. (Yes
(see technical bulletin for specific details for each module)
• Possibility of an integrated barometric reference module, for automatic conversion of pressure measurements
Relative in absolute
• Total possibility of calibrating and configuring Hart instrumentation through a digital communication module
optional integrated
• Total possibility of calibrating and configuring FOUNDATION Fieldbus or Profibus PA instruments, through
integrated optional digital communication module
Full documentation of the tests through integrated data memory, computer interfacability and software of
optional CMX test management and Logical cloud software
• Two simultaneous measurement channels for thermocouple measurements, with universal connection for thermocouples
(connectors/ mini plug) and three channels for temperature resistors (including one for external sample probe)
• Large graphic touch screen display (5.7”) with backlighting and simultaneous display of parameters in
measurement and generation/simulation with multilingual user interface
5 different operating menus on the user interface (depending on the options and modules installed): generator of
temperature, calibrator, calibrator with documentation of the results, datalogger and digital communicator
thorough
• Dimensions and weight: 322x180x298 mm, 9.4 kg
• Power supply: 230V +/ -10%, 50-60 Hz. (the multifunction calibrator section can also be battery-operated)
internal (rechargeable)
• Possibility to combine kits complete with external manual pumps (pneumatic and hydraulic) for the
generation of reference pressures
• Supplied complete power cable, set of test cables and tweezers, USB cable, removal device
inserts and calibration certificate issued by an accredited FINAS laboratory (EA European Accreditation equivalent)
to national standards) and operating manual.
Supplied complete with:
B5671301 - MC6-T150 MH1 — Multihole insert
(3mm, 4mm, 6mm, 8mm, 10mm, 1/4")
B8129900 - Rigid transport case for MC6-T